8.0 TROUBLESHOOTING GUIDE

Symptom/Problem Action/Troubleshooting
Erratic Display

Check electrical connections for continuity and wiring code for pin layout.

No Display/No Output Voltage

Check power connections per instrument Instructions.

Blinking Display

Indicates a signal level greater than full-scale; make sure wires are all
connected, and transducer is within its range.

Incorrect Readout Value

Check transducer range on label. Verify that system was set-up per
instructions. Review set-up procedure. Refer to transducer
troubleshooting guide and verify that transducer operates properly. Use
Shunt Cal to verify calibration.

Limit 1 or 2 fails to trip relay at
setpoint

Verify that limit card is properly installed; check normally open, common,
and normally closed contacts on rear-panel connector for proper wiring.

Peak detector will not reset or will
not hold peak value

Verify that Peak Detector Card is properly installed; check Mode Sel and
common terminal on rear-panel connector for proper wiring.

Track-and-Hold unit will not hold
signal when front switch is pushed

Verify that track-and-hold card is properly installed; check Hold and
Common terminals on rear-panel for proper wiring. Verify that the HOLD
light is coming on in the hold mode.