VersaFluor Filter Selection Guide

Excitation

Catalog

Emission

Catalog

Application

Filters

Number

Filters

Number

DNA Quant (Hoechst 33258)

EX 360/40 (340–380 nm) 170-2420

EM 460/10 (455–465 nm) 170-2421

ß-Galactosidase (MUG)

EX 360/40 (340–380 nm) 170-2420

EM 460/10 (455–465 nm) 170-2421

ß-Glucuronidase (MUGluc)

EX 360/40 (340–380 nm) 170-2420

EM 460/10 (455–465 nm) 170-2421

Apopain (Z-DEVD-AFC)

EX 360/40 (340–380 nm) 170-2420

EM 520/10 (515–525 nm) 170-2424

pH indicator (BCECF)

EX 490/10 (485–495 nm) 170-2422

EM 520/10 (515–525 nm) 170-2424

Calcium indicator (Fluo-3)

EX 490/10 (485–495 nm) 170-2422

EM 520/10 (515–525 nm) 170-2424

DNA Quant (PicoGreen™)

EX 480/20 (470–490 nm) 170-2427

EM 520/10 (515–525 nm) 170-2424

Oligo Quant (OliGreen™)

EX 480/20 (470–490 nm) 170-2427

EM 520/10 (515–525 nm) 170-2424

Protein Quant (NanoOrange™) EX 480/20 (470–490 nm) 170-2427

EM 590/10 (585–595 nm) 170-2425

Fluorescein

EX 490/10 (485–495 nm) 170-2422

EM 520/10 (515–525 nm) 170-2424

Ethidium Bromide

EX 510/10 (505–515 nm) 170-2423

EM 620/10 (615–625 nm) 170-2426

RNA Quant (RiboGreen™)

EX 490/10 (485–495 nm) 170-2422

EM 520/10 (515–525 nm) 170-2424

Helpful Hints

Use filters which are tailored to the sample being measured. Using incorrect
filters will greatly degrade the accuracy of the fluorescence measurement.

• Always use calibrated pipets to insure accurate pipetting.

• Mix standards and sample completely by using a disposable transfer

pipet.

• Remove any air bubbles in the cuvettes.

• Hold fluorometer cuvettes by the upper edges since the cuvettes have

optically clear sides.

• Clean the cuvette sides with a lint-free tissue.

• Read all standards and samples at ambient temperature.

• While taking a reading, keep the samples in the fluorometer only. This

helps reduce sample photobleaching.

• Fluorescence measurements can be made continuously. It is not nec-

essary to zero the instrument before each measurement unless the
baseline conditions change. Check the zero of the instrument with the
baseline fluid (blank) when switching between gain settings.