Fig. 1. Separation of standard.

3

Protein Standard
Components

Component

Molecular Weight

Cytochrome c

12,000

Myoglobin (horse)

17,000

Lysozyme

14,000

Recommended Volume
of Standard

Figure 1 shows separations of the standard using Hi-Pore

®

RP-304 HPLC columns. A volume of 20 µl standard was
applied to each column. The protein peaks give absorbance at
280 nm from about 0.03 to 0.1 absorption units.

2

Conditions
Column:

Reversed Phase
Hi-Pore RP-304, 250 x 4.6 cm

Sample:

20 µl

Eluant:

A. 0.1% TFA
B. 0.1% TFA, 95% acetonitrile
25-50% B linear gradient, 30 min

Flow Rate:

1.5 ml/min

Detection:

UV Monitor @ 280 nm
0.08 AUFS

Peaks:
1. Cytochrome c
2. Myoglobin
3. Lysoz

yme

2

3

1

0

10 20

30

Minutes