beautypg.com

Bio-Rad Bio-Scale™ CHT™ Type I Columns User Manual

Page 9

background image

Fig. 2. Instructions for removal of top frit.

1. Remove the upper and lower end fittings from the column.

2. Remove the upper lock nut from the top end cap.

3. Unscrew the bottom end cap from the protective shield and slide out the glass column.

4. Pull out the upper bed support from the glass column.

5. Using the frit removal tool, remove the distribution screen and frit by pressing the hook

into the frit in a sideways motion with slight downward pressure.

6. Add 5 ml of buffer to the top of the resin bed. Place a new frit and screen into the tube.

Insert the upper bed support and use it to carefully push down the frit and screen until it
just touches the top of the resin bed.

7. Insert the glass column through the protective shield, into the top end-cap and align the

bed support prong with the top end-cap guide. Attach the bottom end-cap to the shield.
Insert the lock nut into the top end-cap and twist down until the plunger is at the top of the
bed.

8. Attach column to a pump and pump buffer at 2.0 ml/min for 5 to 10 minutes. Inspect the

column for voids and adjust accordingly.

7

FLOW

Upper end fitting

Upper lock nut

Top end cap

Protective shield

Upper bed support

Distribution screen

Top frit

Glass column

Bottom end cap

Lower end fitting