Error Message
Probable Cause
Recommended Solution
SCSI ini.
SCSI chip trouble.
Contact Bio-Rad.
WCS R/W SCSI
SCSI chip trouble.
Contact Bio-Rad.
DPR R/W SCSI
SCSI chip damaged.
Contact Bio-Rad.
SCSI bus
Data communication problem
between the Ultramark and
your computer.
The Ultramark is not the source of the
problem. Check your computer.
Bus Rst SCSI
The termination switch may be
off.
Turn the SCSI termination switch on
(located next to SCSI connector).
Another connected SCSI
device is powered off.
Turn the connected machine on, or
turn the termination switch on the
Ultramark on.
See also other documents in the category Bio-Rad Measuring instruments:
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- iMark Microplate Absorbance Reader (112 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)