Bio-Rad Model 550, Ultramark, and Benchmark Microplate Reader Accessories User Manual
Page 22

16
Error Message
Probable Cause
Recommended Solution
Xe flash
The Xenon Lamp did not flash. If this occurs only a few times a year,
it is no problem.
Hi-Voltage power supply
damaged.
Contact Bio-Rad.
Xenon Flash Lamp damaged. Contact Bio-Rad.
Gain Mes-1
Measuring electronics
damaged.
Contact Bio-Rad.
Analog offset voltage is too
high.
Contact Bio-Rad.
Gain Mes-2
Selected filter is not installed. Install
filter.
Measuring electronics
damaged.
Contact Bio-Rad.
Gain Mes-3
Xenon Flash Lamp damaged. Contact Bio-Rad.
Measuring electronics
damaged.
Contact Bio-Rad.
Optical Block is out of the
starting point.
Contact Bio-Rad.
Lenses and Mirrors may be
dirty.
Contact Bio-Rad.
Gain Ref-1
Measuring electronics
damaged.
Contact Bio-Rad.
Analog offset voltage is too
high.
Contact Bio-Rad.
Gain Ref-
Measuring electronics
damaged.
Contact Bio-Rad.
Gain Ref-3
Xenon Flash Lamp damaged. Contact Bio-Rad.
Measuring electronics
damaged.
Contact Bio-Rad.
Lenses and Mirrors may be
dirty.
Contact Bio-Rad.
A/D Mes.
Malfunction of A/D converter. Contact Bio-Rad.
A/D Incu
Malfunction of A/D converter. Contact Bio-Rad.
EEPROM
EEPROM chip trouble.
Contact Bio-Rad.