Error Message

Probable Cause

Recommended Solution

Xe flash

The Xenon Lamp did not flash. If this occurs only a few times a year,

it is no problem.

Hi-Voltage power supply
damaged.

Contact Bio-Rad.

Xenon Flash Lamp damaged. Contact Bio-Rad.

Gain Mes-1

Measuring electronics
damaged.

Contact Bio-Rad.

Analog offset voltage is too
high.

Contact Bio-Rad.

Gain Mes-2

Selected filter is not installed. Install

filter.

Measuring electronics
damaged.

Contact Bio-Rad.

Gain Mes-3

Xenon Flash Lamp damaged. Contact Bio-Rad.

Measuring electronics
damaged.

Contact Bio-Rad.

Optical Block is out of the
starting point.

Contact Bio-Rad.

Lenses and Mirrors may be
dirty.

Contact Bio-Rad.

Gain Ref-1

Measuring electronics
damaged.

Contact Bio-Rad.

Analog offset voltage is too
high.

Contact Bio-Rad.

Gain Ref-

Measuring electronics
damaged.

Contact Bio-Rad.

Gain Ref-3

Xenon Flash Lamp damaged. Contact Bio-Rad.

Measuring electronics
damaged.

Contact Bio-Rad.

Lenses and Mirrors may be
dirty.

Contact Bio-Rad.

A/D Mes.

Malfunction of A/D converter. Contact Bio-Rad.

A/D Incu

Malfunction of A/D converter. Contact Bio-Rad.

EEPROM

EEPROM chip trouble.

Contact Bio-Rad.