xMark Microplate Spectrophotometer User Manual


2

Stability & Drift

≤ 0.010 OD at OD=1 at 490 nm single wavelength reading

Warm-up time

3 minutes

Minimum kinetic read interval

96-well plate:

8 seconds single wavelength

384-well plate: 15 seconds single wavelength
1536-well plate: 35 seconds single wavelength

Shaking

Programmable time period: 0

999 seconds

Shaking mode: Linear and Orbital
Shaking speed: High, medium, low and manual speed

Incubation

25

°C to 45°C, programmable in 0.1°C increments

Set point must be 4

°

C above ambient temperature.

Set point accuracy:

≤ 0.5°C when set point=37°C at RT=25°C

Well-to-well uniformity:

≤ 0.5°C when set point=37°C at RT=25°C

Chamber warm-up time:

≤ 25 minutes when set point=37°C at

RT=25

°C

Imaging

Maximum imaging area: 124 x 82 mm
Resolution: 0.50, 0.75, 1.00 mm per defined wavelength

Wavelength scanning

Range: 200–1000 nm, Minimum increment: 1 nm per defined plate well

Environmental operating range

10–40ºC, max. relative humidity 95% noncondensing

Power

100–240 V, 60/50 Hz, 350 VA max

Dimensions

544 mm (W) x 440 mm (D) x 219 mm (H), including feet

Weight

27.5 Kg, including incubator

Computer interface

USB 2.0 communication port

Software compatibility

Windows XP, Vista, Mac OS 10.X