13. After the detection antibody incubation, slowly remove and discard

the sealing tape, then

vacuum-filter.

14.

Wash 3 times.

15. Vortex the streptavidin-PE (1x) vigorously and add 50 µl to each well.

Incubate for 10 min.

16. After the streptavidin-PE incubation, slowly remove and discard the

sealing tape, then

vacuum-filter.

17.

Wash 3 times.

18. Add 125 µl of assay buffer to each well. Cover the filter plate with a

new sheet of sealing tape. Shake the filter plate at room temperature
at 1,100 rpm for 30 sec and remove the sealing tape before
reading the plate.

# of Wells

100x

Streptavidin-PE

(µl)

Total Volume (µl)

96

48

32

24

60

30

20

15

5,940

2,970

1,980

1,485

6,000

3,000

2,000

1,500

Assay Buffer (µl)

Example Streptavidin-PE Calculations