B. dilution of standard (1:3 serial dilution), C. sample preparation – Bio-Rad Apoptosis Assays User Manual
Page 2

B. Dilution of Standard (1:3 Serial Dilution)
1. Label 8 polypropylene tubes S1 through S8.
2. Transfer the reconstituted standard into the tube labeled “S1.”
3. Add the appropriate amount of the standard diluent into the labeled
tubes according to the table below (this will be sufficient for duplicate
standard curves and blanks).
4. Prepare working standards (S2–S8) by serial dilution. Transfer the
appropriate volume of standard into each of the labeled tubes with
standard diluent as outlined above.
5. Vortex each standard at a medium setting before proceeding with the next
serial dilution. Change pipet tip at each dilution step.
C. Sample Preparation
1. Refer to instruction manual #10033631 for detailed sample preparation.
Note: Pay close attention to cell lysis, homogenization, and fractionation
protocols in the manual. Analytes of interest may be localized to the
cytosolic fraction or to the nuclear and mitochondrial fraction.
2. Prepare sample dilutions in 0.5 ml or 1.0 ml polypropylene tubes as
required for the assay.
Bio-Plex Pro Assay Quick Guide
Standard
Volume of Standard Diluent, µl
Volume of Standard, µl
S2
100
50 of S1
S3
100
50 of S2
S4
100
50 of S3
S5
100
50 of S4
S6
100
50 of S5
S7
100
50 of S6
S8
100
50 of S7
Blank
100
—