Initial Preparation
1. Plan the plate layout.
2. Start up/warm up the Bio-Plex
®
system (30 min).
n
Bring diluents including wash buffer, assay buffer, standard diluent
HB, detection antibody diluent HB, and sample diluent HB to room
temperature (RT). Keep other items on ice until needed
n
Begin to thaw frozen samples
3. Prime wash station for flat bottom plate or set vacuum manifold to
–1 to –3" Hg for filter plate.
4. Calibrate the Bio-Plex system by following the prompts within the
Bio-Plex Manager
™
software. This can be done now or during an
assay incubation step.
5. Prepare 1x wash buffer. Mix 10x stock by inversion to ensure all salts
are in solution. Then dilute 1 part 10x wash buffer with 9 parts dH
2
0.
6. Prepare 10% serum-based standard diluent by diluting 1 part standard
diluent HB to 1.5 parts sample diluent HB (4 ml standard diluent HB to
6 ml sample diluent HB).