Page 13
21
Troubleshooting Guide
(continued)
Probable
Recommended
Problem
Cause
Solution
e. Primary antibody
i. Loss of reac
is not specific or
tivity may
does not recognize
have occurred
denatured antigens
during elec-
(common with
trophoretic
monoclonals).
transfer. Pre-
test the reactiv-
ity of the anti-
body against
the antigen by
a dot blot.
2. High background. a. Exposure time was
i. Decrease
too long.
exposure time.
b. PBST washes after
i. The washes are
transfer were omit-
critical to
ted or insufficient.
reduce spotted
or blotchy
background
development.
c. Blocking was insuf-
i. Increase the
ficient.
time of the
blocking step
and/or the con-
centration of
blocker used.
20
Troubleshooting Guide
(continued)
Probable
Recommended
Problem
Cause
Solution
d. Wash stringency
i. Tween-20 is
was insufficient.
necessary in
wash steps to
reduce back-
ground. The
concentration
can be
increased up to
0.3% if back-
ground persists.
ii. Increase the
number and
length of
washes.
e. Second antibody
i. Use the recom-
conjugate was used
mended dilu
at an excessive con-
tion,or deter-
centration.
mine the opti-
mal dilution
experimentally.
f. Contamination
i. Refer to the
occurred during
instrument
transfer.
instruction
manual for rec-
ommendations.
g. Insufficient number
i. Increase
of DMSO/PBST
number of
washes.
wash steps
following
BAR.
4100130c.qxd 4/9/99 8:29 AM Page 20