21

Troubleshooting Guide

(continued)

Probable

Recommended

Problem

Cause

Solution

e. Primary antibody

i. Loss of reac

is not specific or

tivity may

does not recognize

have occurred

denatured antigens

during elec-

(common with

trophoretic

monoclonals).

transfer. Pre-
test the reactiv-
ity of the anti-
body against
the antigen by
a dot blot.

2. High background. a. Exposure time was

i. Decrease

too long.

exposure time.

b. PBST washes after

i. The washes are

transfer were omit-

critical to

ted or insufficient.

reduce spotted
or blotchy
background
development.

c. Blocking was insuf-

i. Increase the

ficient.

time of the
blocking step
and/or the con-
centration of
blocker used.

20

Troubleshooting Guide

(continued)

Probable

Recommended

Problem

Cause

Solution

d. Wash stringency

i. Tween-20 is

was insufficient.

necessary in
wash steps to
reduce back-
ground. The
concentration
can be
increased up to
0.3% if back-
ground persists.

ii. Increase the

number and
length of
washes.

e. Second antibody

i. Use the recom-

conjugate was used

mended dilu

at an excessive con-

tion,or deter-

centration.

mine the opti-
mal dilution
experimentally.

f. Contamination

i. Refer to the

occurred during

instrument

transfer.

instruction
manual for rec-
ommendations.

g. Insufficient number

i. Increase

of DMSO/PBST

number of

washes.

wash steps
following
BAR.