Appendix F
Technical Specifications
Supply Voltage:
115 to 220 VAC - 50 to 60 Hz
Current Draw:
1 Amp. Max.
Fuse (1/4 * 1-1/4):
Fast Blow 1 Amp.
Cable Requirements
AVC Display Module to Remote Sensor:
Distance:
Maximum 2,500 feet (750 m)
Cable:
Belden # 8451 or equivalent.
#22 gauge shielded, twisted pair (For multiple sensors a multi-conductor shielded
twisted pair cable should be used).
Signal:
30
VDC
- 1.0
Amps.
60
VDC
- 0.3
Amps.
125 VAC
-
0.5 Amps.
AVC Display Module to Master Module
Distance:
Maximum 10,000 feet (3,000 m)
Cable:
Minimum #22 gauge wire (does not have to be shielded, twisted pair)
Signal:
5 VDC, < 5 µA
www.amico.com
31
This manual is related to the following products:
See also other documents in the category Amico Equipment:
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Motorized Retractable (20 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alert-3 LCD Alarm (28 pages)
- Automatic Cryogenic Liquid Solenoid Manifold (32 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- British Standard Outlet Slim (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold CSA (44 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)