Riser plate to outlet detail c - riser, Approval signature date phone no – Amico Ceiling Column Manual Retractable User Manual
Page 9
Approval Signature
Date
Phone No.
Access Panel This Side
NOTE:
1.
Riser plate is completed with DISS pipe stubs, typical for each gas (For DISS riser pipe inlet see page 11).
2.
Pipe stubs diameters are 3/8” [9.97 mm] ID and 1/2” [12.7 mm] OD for all gases. Refer to Riser Pipe cut sheet.
3.
Exact locations of gas terminations on riser plate will be determined after receiving this booklet.
BOTTOM VIEW COLUMN
Looking Up
Access Panel This Side
X4 See
page 9
2A
1A
Access panel This side
Side A
B
D
C
4.5” [114.3 mm] above
7/8” [22.2mm] hole X8
Termination X4
box use
see page 10
Mounting holes for
threaded rod X4
see note #4
for column X4
Access Panel This Side
Riser Plate Bottom View Looking Up
pipe stub
fi nished ceiling
For DISS riser pipe
inlet (see page 9)
anti-sway support
(supplied by others)
17”
[431.8 mm]
14.75”
[374.65 mm]
1.00”
[25.4 mm]
1.00”
[25.4 mm]
14.75”
[374.65 mm]
anti-sway support
(supplied by others)
for electrical conduit use
DISS
14.75”
[374.65 mm]
17”
[431.8 mm]
BOTTOM VIEW
Riser Plate to Outlet Detail C - Riser
TOP VIEW
RISER PLATE
www.amico.com
7