Electrical diagram (bottom view) – Amico Ceiling Column Manual Retractable User Manual
Page 8
*Please Select One of the Following:
Amico recommended circuit schedule based on layout
Custom circuit schedule. Please specify on circuit chart
Receptacle Type:
US/CAN
*BRITISH
(simplex only)
*GERMAN
(simplex only)
Circuit Chart
Electrical Diagram (Bottom View)
Approval Signature
Date
Phone No.
NOTE:
1.
Two electrical/communication/medical gas devices
cannot be placed together at any corner.
2.
If type of wire & gauge is not specified, Amico
will use the following:
For US/International
LIVE/HOT - 12 Ga. THHN Wire Black
NEUTRAL - 12 Ga. THHN Wire White
GROUND - 10 Ga. THHN Wire Green
For Canada
LIVE/HOT - 12 Ga. RW90 Wire Black
NEUTRAL - 12 Ga. RW90 Wire White
GROUND - 10 Ga. RW90 Wire Green
3.
Conduit: 9/16” [14.28 mm] fl exible metal conduit.
4.
All conduits and wires will extend a minimum of 12”
[304.8 mm] above fi nish column for connection
to riser plate.
5.
A pull-string will be provided to be connected by
others for all ground jacks and data provision.
6.
British or German Receptacle cannot be positioned
side by side i.e.: A1 and A2.
*Please specify electrical details:
Type of wire:
THHN
XHHW
RW90
AWG Wire Gauge:
Wire Color:
10 & 12 LIVE/HOT
10 & 12 NEUTRAL
10 GROUND
Isolated
Power
Please select one of the following
All receptacles shall be mounted with
ground
pin
“DOWN”
All receptacles shall be mounted with
ground pin “UP” (Default)
6
Amico Pipeline
____ gauge
____ gauge
____________________
____________________
10 gauge
Green