background image

Servicing

Approval Signature

Date

Phone No.

The periodic inspection of gas outlets and electrical outlet is recommended to ensure proper safety and operation.

A) Gas Outlet

* Inspect the gas outlet connection points for the following:

1. Proper and legible identifi cation.

2. Worn or defective engagement mechanisms.

3.

Leaks.

4. Deformed or damaged threaded connections.

5. Loose fasteners.

B) Electrical

* Inspect the electrical outlet connection points for the following:

1. Proper and legible identifi cation.

2. Deformation or damage

3.

Discoloration

WARNING

Always shut off source pressure at designated zone valves before servicing any gas outlet.

WARNING

Use care when wiping down electrical outlets to prevent electrical shock.