Monarch Instrument TACH-4AR User Manual

Page 6

background image

POWER/LOCK

When the instrument is “locked on”, the POWER

button will turn the unit off. When not “locked

on” the POWER button must be depressed and

held to take a reading. Each time POWER is

pressed, the previous Minimum and Maximum

value is cleared, and the optics are turned on to

take the next reading. When the POWER button is

released, the Tachometer holds the last reading for

60 seconds, then automatically shuts off.

NOTE: None of the following buttons will function when the POWER

button is being held.

RECALL/STORE

Used to display the stored Maximum or Minimum

value. When pressed, the unit will display “MAX”

or “MIN”, then the value. The reading alternates

between the Maximum and Minimum values each

time this button is pressed. The value is displayed

for 2 seconds, then the display returns to the last

measurement.

RESET/LOCK

Resets the minimum, maximum, and last measure

ment to Zero.

LEFT ARROW (")

Turns the LCD back light on for 60 seconds

(assuming the backlight is not already on)

SCALE MODE

Operation is the same as for the RPM mode, except that the measurement

is scaled. In the RPM mode, the input frequency is automatically

multiplied by 60 to display in Revolutions Per Minute. In the scale mode,

the input frequency (pulses per second) is multiplied by a user defined

scale factor. As an example: for two pulses per revolution and a desired

readout in RPM, the Scale Factor is 30. (60 / pulses per rev = Scale

Factor) The scale factor can range from 0.0001 to 99999. This allows the

user to display data in real engineering units.

REMOTE SENSORS (TACH-4AR ONLY)

The remote sensors plug into the stereo phone socket on the end panel

of the unit . The unit automatically detects the remote sensor at ‘turn on’

time and powers the remote sensor on as needed. If you plug a remote

sensor in while in the middle of a measuring operation, the unit will not

recognize the remote sensor. It is then necessary to turn the TACH-4A

off, then on again. Remember to press and hold the POWER button

or ‘lock-on’ the unit to take a reading with a remote sensor.

It is often useful to turn the TACH-4A on in the “LOCK” mode when using

remote sensors. Plug the sensor in, then press and hold the RESET/

LOCK button and press the POWER button to turn the unit on and

permanently power the sensor. Refer to individual modes for further

instruction.

PO, PULSE OUTPUT (OPTIONAL)

The Pulse Output is an option for both the TACH-4A and TACH-4AR. This

option includes a 2.5mm phone socket installed on the lens panel of the

Tachometer and a six foot cable with mating 2.5mm phone plug to male

BNC connector enabling the user to interface with a standard BNC cable.

The Pulse Output is a 0 to 5 Volt TTL compatible output. Every input

pulse gives an output pulse. Normally, the Tachometer is used with one

pulse per revolution, therefore the output would be one pulse per revolution.

If the Tachometer is used with more than one pulse per revolution, the

output will echo the input, thus providing more than one pulse per

revolution.