Monarch Instrument TACH-4AR User Manual

Page 4

background image

OPERATING DISTANCE AND ANGLE

Measurements can be made up to 36 inches and at angles of up to 45

°

from

the reflective tape (T-5). When measuring shiny surfaces, a 20

°

angle is

recommended. The TACH-4A has exceptional accuracy, and for very low or

very high speed measurements, it is desirable to hold the instrument against

a steady rest, or mount it on a Tripod or bracket using the 1/4 - 20 UNC

mounting bushing on the back side of the instrument. This ensures that

slight hand motion will not influence the time interval between pulses,

giving erroneous readings. For most measurements, it is not necessary to

mount the instrument in this manner. The TACH-4AR with RLS-5P Laser

Sensor can be used up to 50 feet and 45

°

from the reflective tape (T-5).

CHANGING THE MODE

To change the operating mode of the unit, press the MENU button so that

“MENU” is displayed. Pressing the UP ARROW (!) button once will

display the current mode. Keep pressing the UP ARROW (!) button until

the desired mode is displayed. Press the LEFT ARROW (") button to

change an option of that mode, i.e. to select MANUAL or AUTO for the

Timer mode. Finally, press the RECALL (STORE) button to “save” the

new mode. To set the scale factor for Scale or Totalize modes see Setting

the Scale Factor section.

MENU SELECTIONS

Pressing the MENU button causes the word “Menu” to be displayed. Pressing

the UP ARROW (!) button displays the current mode. Pressing the UP

ARROW (!) button again displays the RPM mode. Press the UP

ARROW (!) button again to display the next mode. TEST loops around

back to RPM. Pressing the LEFT ARROW (") button will perform the

function stated in the following table or display the current option setting.

Pressing the LEFT ARROW (") button again allows selection of either of

the two options shown.

Do not forget to Press the RECALL (STORE) button to save the selected

mode or value. Pressing the MENU button returns to the “MENU”

function. Pressing the RESET button returns to the operating mode

without changing the unit.

ACCESSORIES

MODEL

DESCRIPTION

ROS-5P

Remote Optical Sensor with an 8 foot cable and

a mounting bracket

RLS-5P

Remote Laser Sensor with an 8 foot cable and a

mounting bushing on a base Infrared Sensor with

an 8 foot cable

IRS-5P

Infrared sensor with an 8 foot cable

MT-190P

Magnetic Trigger Sensor/Amplifier with an 8 foot

cable

EC-25P

25 foot extension cable for remote sensors with

male/female 1/8 inch phone plug connectors

R-5B

Recharger, 115 Vac - 50/60 Hz, with 4 “AA” Nicad

Batteries

R-6B

Recharger, 230Vac - 50/60 Hz, with 4 “AA” Nicad

Batteries

B-4

Four “AA” Rechargeble NiCad Batteries

CC-3

Plastic Latching Carrying Case

T-5

Reflective Tape - 5 foot roll, 0.5 inch wide

T-5WP

Waterproof Reflective Tape (honeycomb pattern)

5 foot roll, 1 inch wide

CAL-N.I.S.T

N.I.S.T. Traceable Certificate of Calibration and

Re-Calibration