LaMotte PHOSPHONATE TEST KIT FI-DR User Manual
L mott, Phosphonate test kit
PHOSPHONATE TEST KIT
DIRECT READING TITRATOR METHOD
MODEL FI-DR • CODE 7530-DR
QUANTITY
CONTENTS
CODE
15 mL
*Hydrochloric Acid, 1.0N
*6130-E
15 mL
*Fluoride Inhibitor
*3929-E
15 mL
Sodium Thiosulfate, 0.1N
6155-E
10 g
Xylenol Orange Powder
6165-D
60 mL
Thorium Nitrate Solution
6158PS-H
1
Test Tube, 5-10-15 mL, w/cap
0778
1
Dispenser Cap
0601
1
Spoon, 0.1 g
0699
1
Direct Reading Titrator, 0-20 Range
0378
1
pH paper, 1.3-4.4
2958
*WARNING:
Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these
reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
number.
Carefully read the instruction manual for the Direct Reading Titrator before
performing the titration procedure described below. The Titrator is calibrated in
terms of parts per million Phosphonate, and each minor division on the Titrator
scale equals 0.4 ppm.
NOTE: This test has been calibrated for Dequest 2006. When a different
compound is to be tested, the amount of Thorium Nitrate added should be
multiplied by a conversion factors below to determine ppm Phosphonate. If any
other phosphonate compound is used, the factor must be determined
experimentally using standard solutions of that compound.
Phosphonate
Compound Name
Factor
Dequest 2000
AMP(NTP)
0.8
Dequest 2006
NaAMP
1.0
Dequest 2010
HEDP(A)
0.6
Belcor 575
0.5
L Mott