LaMotte PERACETIC ACID TEST KIT 7191-01-FMC User Manual
Peracetic acid test kit
PERACETIC ACID TEST KIT
CODE 7191-01-FMC
QUANTITY
CONTENTS
CODE
30 mL
*Sulfuric Acid, 1:1
*6141WT-G
15 mL
Ferroin Indicator
6410-E
2 x 30 mL
*Hydrogen Peroxide Titrant
*5650LWT-G
30 mL
*Potassium Iodide 20% Solution
*6521-G
60 mL
Peracetic Acid Titrant
S-6155-H
1
Test Tube, 5-10-25 mL, plastic, w/cap 0715
1
Pipet, 0.5 mL, plastic
0353
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these
reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
number.
NOTE:
Do not use for tapwater or less than 20 ppm peracetic acid.
PRO CE DURE
CONCENTRATION OF HYDROGEN PEROXIDE
1.
Fill test tube (0715) to 10 mL line with sample water.
2.
Add 10 drops of *Sulfuric Acid, 1:1 (6141WT). Swirl to mix. Make
sure solution is thoroughly mixed.
3.
Add 1 drop of Ferroin Indicator (6410). Swirl to mix. Make sure
solution is thoroughly mixed. Solution will turn pale orange-pink if
hydrogen peroxide is present.
4.
While gently swirling tube, add *Hydrogen Peroxide Titrant
(5650LWT) one drop at a time until pale orange-pink color changes
to colorless or pale yellow. Make sure solution is thoroughly mixed.
Count the number of drops added. Hold bottle vertically.
5.
Discard titrated sample and rinse tube out with sample water.
6.
Multiply the number of drops used in Step 4 by 50 (fifty). Record as
ppm Hydrogen Peroxide.