beautypg.com

Lamotte com pany – LaMotte CALCIUM AND TOTAL HARDNESS IN WATER TEST KIT User Manual

Page 2

background image

4.

While swirling tube, add Hardness Titrant 10 (2783WT) one drop at
a time, until color changes to blue. Count the number of drops added.
Hold bottle vertically.

5.

Multiply number of drops used in Step 4 as follows:

25 mL

multiply by 10

10 mL

multiply by 25

5 mL

multiply by 50

Record the result as ppm Total Hardness as CaCO

3

.

CAL CIUM HARD NESS
1.

Rinse test tube with sample water. Fill with desired sample size selectd
from table above.

2.

Add *Sodium Hydroxide with Metal Inhibitor (4259) as follows:
25mL

12 drops

10 mL

5 drops

5 mL

3 drops

3.

Add one Calcium Hardness Indicator Tablet (T-5250). Swirl to mix.
Solution should be red.

4.

While swirling tube, add Hardness Titrant 10 (2783WT) one drop at
a time, until color changes to blue. Count the number of drops added.
Hold bottle vertically.

5.

Multiply number of drops used in Step 4 as follows:

25 mL

multiply by 10

10 mL

multiply by 25

5 mL

multiply by 50

Record the result as ppm Calcium Hardness as CaCO

3

To determine Magnesium Hardness, subtract Calcium Hardness from Total
Hardness.

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Challenges

SM

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

10/08