4.
While swirling tube, add Hardness Titrant 10 (2783WT) one drop at
a time, until color changes to blue. Count the number of drops added.
Hold bottle vertically.
5.
Multiply number of drops used in Step 4 as follows:
25 mL
multiply by 10
10 mL
multiply by 25
5 mL
multiply by 50
Record the result as ppm Total Hardness as CaCO
3
.
CAL CIUM HARD NESS
1.
Rinse test tube with sample water. Fill with desired sample size selectd
from table above.
2.
Add *Sodium Hydroxide with Metal Inhibitor (4259) as follows:
25mL
12 drops
10 mL
5 drops
5 mL
3 drops
3.
Add one Calcium Hardness Indicator Tablet (T-5250). Swirl to mix.
Solution should be red.
4.
While swirling tube, add Hardness Titrant 10 (2783WT) one drop at
a time, until color changes to blue. Count the number of drops added.
Hold bottle vertically.
5.
Multiply number of drops used in Step 4 as follows:
25 mL
multiply by 10
10 mL
multiply by 25
5 mL
multiply by 50
Record the result as ppm Calcium Hardness as CaCO
3
To determine Magnesium Hardness, subtract Calcium Hardness from Total
Hardness.