PRO CE DURE
HIGH RANGE: 10-100 PPM PHOS PHATE (PO
4
)
1:10 DI LU TION
1.
Use the 1.0 mL pipet (0354) to add 1.0 mL of sample water to the
test tube (0970-S).
2.
Dilute to 10 mL line with distilled or phosphate free water.
3.
Use the 0.5 mL pipet (0353) to carefully add 0.5 mL of *Phosphate
Reagent A (4606).
4.
Add one *Phosphate B/C Tablet (4607). Cap and gently mix until
the tablet is dissolved. Wait 4 minutes.
5.
Fill a 5 mL test tube (0230) to the 5 mL line with this reacted
sample.
6.
Insert test tube into Phosphate Comparator (7899). Match sample
color to a color standard. Record as ppm Phosphate.
NOTE:
If sample color is lighter than the lightest color standard,
use the following procedure.
LOW RANGE: 1-10 PPM PHOS PHATE (PO
4
)
UN DI LUTED SAM PLE
1.
Fill the test tube (0970-S) to the 10 mL line with sample water.
2.
Use the 0.5 mL pipet (0353) to carefully add 0.5 mL of *Phosphate
Reagent A (4606).
3.
Add one *Phosphate B/C Tablet (4607). Cap and gently mix until
the tablet is dissolved. Wait 4 minutes.
4.
Fill a 5 mL test tube (0230) to the 5 mL line with this reacted
sample.
5.
Insert test tube into Phosphate Comparator (7899). Match sample
color to a color standard.
IMPORTANT:
Divide the value obtained from the comparator by
10. Record as ppm Phosphate.
CAUTION:
Do not use soaps or detergents containing phosphate to
wash test tubes.
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Chal lenges
®
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com