PRO CE DURE

FREE AVAILABLE CHLORINE
1.

Rinse a test tube (0106) with sample water. Fill test tube to 5 mL line
with sample water.

2.

Add one *Chlorine DPD #1R Tablet (6999A). Cap and shake until
tablet disintegrates.

3.

Immediately insert test tube into Chlorine DPD Comparator (3596).
Match sample color to a color standard. Record as ppm free available
chlorine.
NOTE:

Save sample if total residual and combined chlorine are to be

determined.

TOTAL RESIDUAL CHLORINE & COMBINED CHLORINE
5.

Add one *Chlorine DPD #3R Tablet (6905A) to the sample from
Step 3. Cap and shake until tablet disintegrates.

6.

Insert test tube into Chlorine DPD Comparator (3596). Match
sample color to a color standard. Record as ppm total residual
chlorine.

7.

Calculate combined chlorine:

Combined Chlorine (ppm) =

Total Residual Chlorine - Free Available Chlorine

NOTE:

Thoroughly clean and rinse test tubes after each test.

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Challenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

9/06