beautypg.com

Lamotte com pany – LaMotte DEHA TEST KIT 4790 User Manual

Page 2

background image

3.

Thoroughly clean tubes immediately after each test to avoid
staining and future contamination.

4.

Ferrous ions, Fe

+2

, will give a false positive. The contribution of Fe

+2

to reading can be determined by testing a second sample and
skipping Step 4. This result should be subtracted from previous
result.

PRO CE DURE
1.

Fill test tube (0106) to the 5 mL line with sample water.

2.

Add 3 drops of *DEHA Reagent 1 (4791) and 3 drops of *DEHA
Reagent 2 (4792).

3.

Cap and mix.

4.

Add 3 drops of *DEHA Reagent 3 (4793). Cap and mix.
NOTE:

If the alkalinity of the sample is over 300 ppm, use the

following number of drops of *DEHA Reagent 3:

Alkalinity, ppm

Drops of

*DEHA Reagent 3

300-399

4

400-499

5

500-599

6

600-699

7

700-799

8

800-899

9

900-999

10

5.

Wait exactly 10 minutes.

6.

Immediately insert test tube into top of Octa-Slide Viewer (1100).
Slide DEHA Octa-Slide Bar (3442) into Octa-Slide Viewer. Match
sample color to a color standard. Record result as ppm DEHA.

NOTE:

This test kit will also test for the oxygen scavenger erythorbic

acid (also called isoascorbic acid or araboascorbic acid). Follow the test
instruction and multiply the results by 4 to get ppm Erythorbic Acid.

DEHA x 4 = Erythorbic Acid

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Challenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

1/11