LaMotte HYDROPONICS TEST KIT HP-1 User Manual
Page 2
WARNING: Reagents marked with a * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
see MSDS CD or www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
number.
Read Demineralizer Bottle instruction manual before proceeding.
The Demineralizer Bottle (1152) will be the source of all deionized water
used in these tests.
DI LU TION PRO CE DURE
If the test result in the nitrate, phosphate or potassium test is greater than
the highest value it is necessary to perform a dilution. Use the 1.0 mL
pipet (0354) and the 25 mL graduated cylinder (0417) as follows:
AMOUNT
OF SAMPLE
DILUTE
TO
MULTIPLY
RESULT BY
1.0 mL
5.0 mL
5
1.0 mL
10.0 mL
10
1.0 mL
20.0 mL
20
1.0 mL
25.0 mL
25
Use deionized water from the Demineralizer Bottle for all dilutions.
pH
A pH value of 7.0 indicates a neutral solution. Values above 7.0 indicate
alkaline solutions, while values below pH 7.0 indicate acid solutions.
1.
Fill a test tube (0106) to 5.0 mL line with sample solution.
2.
Add 8 drops of *Wide Range Indicator (2218). Cap and mix.
3.
Insert pH Wide Range Octa-Slide (3424) into Octa-Slide Viewer.
4.
Insert test tube into the Octa-Slide Viewer (1100). Match sample
color to the pH Wide Range Octa-Slide Bar (3424).
NOTE: If the pH of the sam ple is not be tween 6.2 and 6.8, it must be
ad justed to 6.5 be fore pro ceed ing the with the Ni trate, Phos pho rus,
and Po tas sium tests. If the pH is be low 6.2 add *So dium Hy drox ide,
1.0N (4004), one drop at a time, pe ri od i cally check ing the pH un til it
is brought into the range of 6.2 to 6.8. If the pH is above 6.8, add
*Hy dro chlo ric Acid, 1.0N (6130) in the same man ner.