LaMotte AQ-5 Saltwater Aquaculture Kit User Manual
Salt water aquaculture test kit
SALT WATER AQUACULTURE
TEST KIT
MODEL AQ-5 · CODE 3636-03
NOTE: It is important to read the instruction manual
before attempting to perform the tests with the short
form instructions provided below.
*WARNING:
Reagents marked with a * are
considered to be potential health hazards. To view or
print a Material Safety Data Sheet (MSDS) for these
reagents see MSDS CD or www.lamotte.com. To obtain
a printed copy, contact LaMotte by e-mail, phone or
fax.
To order individual reagents or test kit components, use
the specified code number.
AM MO NIA NI TRO GEN
1.
Fill test tube (0106) to 5 mL line with sample water.
2.
Add 10 drops *Salicylate Ammonia #1 (3978LWT).
Cap and mix.
3.
Add 7 drops of *Salicylate Ammonia #2 (3979WT).
Cap and mix. Wait 1 minute.
4.
Add 7 drops of Salicylate Ammonia #3 (3982WT). Cap
and mix. Wait 20 minutes.
5.
Insert test tube into Octa-Slide Viewer (1100) with
Ammonia Nitrogen Octa-Slide Bar (3438) inserted.
6.
Record as ppm Ammonia Nitrogen (NH
3
-N).
NI TRITE NI TRO GEN
1.
Fill test tube (0106) to 2.5 mL line with sample water.
2.
Dilute to 5 mL line with *Mixed Acid Reagent
(V-6278).
3.
Use 0.1g spoon (0699) to add 0.1g of *Color
Developing Reagent (V-6281). Cap and mix for 1
minute. Wait 5 minutes.
4.
Insert test tube into Octa-Slide Viewer (1100) with
Nitrite Nitrogen Octa-Slide Bar (3437) inserted.
5.
Record as ppm Nitrite Nitrogen (NO
2
-N).
NI TRATE NI TRO GEN
1.
Fill test tube (0106) to 2.5 mL line with sample water.
2.
Dilute to 5 mL line with *Mixed Acid Reagent
(V-6278). Cap and mix. Wait 2 minutes.
3.
Use 0.1g spoon (0699) to add one level measure of
*Nitrate Reducing Reagent (V-6279). Cap and invert
tube 50-60 times in one minute. Wait 10 minutes.
4.
Invert sample one time and insert test tube into
Octa-Slide Viewer (1100) with Nitrate Nitrogen
Octa-Slide Bar (3413) inserted.
5.
Record as ppm Nitrate Nitrogen (NO
3
-N).
AL KA LIN ITY
1.
Fill titration tube (0608) to 5 mL line with sample
water.
2.
Add 4 drops of *BCG-MR Indicator (2311-EG). Cap
and mix. Sample will turn blue-green.
3.
Fill Direct Reading Titrator (0382) with *Alkalinity
Titration Reagent B (4493DR).
4.
Titrate sample until blue-green color changes to pink.
5.
Record as ppm Alkalinity (CaCO
3
).
CARBON DI OX IDE
1.
Fill titration tube (0608) to 20 mL line with sample
water.
2.
Add 2 drops *Phenolphthalein Indicator, 1% (2246). If
sample turns red, no free carbon dioxide is present. If
colorless, proceed to Step 3.
3.
Fill Direct Reading Titrator (0380) with *Carbon
Dioxide Reagent B (4253DR).
4.
Titrate sample until faint pink color persists for 30
seconds.
5.
Record as ppm Carbon Dioxide (CO
2
).
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
SM
PO Box 329 · Chestertown · Mary land · 21620 · USA
800-344-3100 · 410-778-3100 (Out side U.S.A.) · Fax 410-778-6394 · www.lamotte.com