beautypg.com

LaMotte AR-42 User Manual

Page 4

background image

4

7.

TOTAL HARDNESS TEST PROCEDURE

Add 1 Hardness

Reagent #6 Tablet

(4484).

Swirl to mix.

While gently swirling the tube, add

Hardness Reagent #7 (4487WT)

one drop at a time until the red

color changes to blue.

Count the number of

drops added.

Hold bottle

vertically.

Multiply the number of drops used in Step 6 as follows:
upper line:

multiply by 10 = ppm Hardness as CaCO

middle line:

multiply by 1 = gpg Hardness as CaCO

lower line:

multiply by 20 = ppm Hardness as CaCO

3

3

3

Fill the test tube (4488) to the desired line with the

sample water.
upper line:

1 drop = 10 ppm CaCO

middle line:

1 drop = 1 gpg CaCO

lower line:

1 drop = 20 ppm CaCO .

3

3

3

Add 5 drops

of *Hardness

Reagent #5

(4483).

1.

10
1

20

10

1

20

10
1

20

10
1

20

10

1

20

Swirl until

tablet has

disintegrated.

Solution will

turn red.

10

1

20

2.

3.

4.

5.

6.