LaMotte pH PLUS Direct Meter User Manual
Page 18

5. To cancel or abort this operation, press the
CAL key. Note no new value
will be stored into the meter's non-volatile memory. To confirm the
calibration, press the
ENTER key. The LCD displays “CO” momentarily,
and the meter reverts to the measurement mode.
3.4.2 Without Temperature probe (no ATC)
If a temperature probe is not used, the meter compensates for the pH response
based on a new calibrated temperature value manually set or at 25.0 °C
(factory default).
1. Press the
MODE key to enter into Temperature mode until “°C” shows on
the LCD.
2. Compare the displayed value to a NIST certified thermometer or
thermometer known to be accurate (dipped into a constant temperature
bath).
3. Press the
CAL key to enter the temperature calibration mode. The LCD
shows “
CA” momentarily and the displayed reading flashes. Note that this
displayed value should either be 25.0
o
C or the last set temperature value.
4. Press
p and q keys until the display shows the desired temperature. Any
value from 0 to 100
o
C can be set.
5. To cancel or abort this operation, press the
CAL key. Note no new value
will be stored into the meter‘s non-volatile memory. To confirm the
calibration, press the
ENTER key. The LCD displays “CO” momentarily,
and the meter reverts to measurement mode.
18