beautypg.com

LaMotte DO 4000 Dissolved Oxygen Meter User Manual

Page 17

background image

PERCENT SATURATION READING
1. Press “%SAT” button to turn meter on.
2. Set the Salinity Compensation knob to the salinity of the sample

water.

3. Keeping probe at an angle, slowly lower at least 3.5” into sample.

Tap probe against the side of the beaker to remove air bubbles. Stir

solution gently with probe until display stabilizes. If testing deeper

water, gently raise and lower probe several times to remove air

bubbles. Continue to raise and lower to avoid oxygen starvation at

the membrane.
If readings are erratic see Troubleshooting on page 18.

4. Record reading as percent (%) Saturation.
5. Press “OFF” button and turn off constant polarization unless meter

is to be used again shortly.

17