CALIBRATING
The CDS 5000 is precalibrated at the factory; set both the TDS knob and
the SLOPE knob to the 12 o’clock position and proceed. This procedure
will give results within

±10% of the actual reading. For more accurate

results, follow the procedure below.
1. Press “°C” button to turn the meter on.
2. Set TDS knob to the 12 o’clock position.
3. Insert temperature and conductivity probes at least

1

2

“ into

conductivity standard.

4. Gently stir with conductivity probe until reading stabilizes. Press

mS/ppm” or “mS/ppt” button.

5. Adjust SLOPE knob until display reads conductivity of chosen

standard. The CDS 5000 is now calibrated and ready for use.

LaMotte offers several ranges of conductivity standards. Choose the
standard most appropriate for your testing needs and order using the four
digit code number listed.
STANDARDS

Description

Code #

Cond. Std., 0.0005M KCl, 74

mmhos/cm

6416-L

Cond. Std., 0.005M KCl, 718

mmhos/cm

6417-L

Cond. Std., 0.01M KCl, 1,413

mmhos/cm

6354-L

Cond. Std.,0.05M KCl, 6,668

mmhos/cm

6418-L

Cond. Std., 0.5M KCl, 58,640

mmhos/cm

6419-L