5. ELECTRODE CARE AND MAINTENANCE

For best results, store the pH probe bulb wet. Store the pH glass bulb
in pH probe storage solution. NEVER use deionized water for storage.
Wash probes with clean water after each use.

pH probes are susceptible to contamination and dirt. Clean them as
needed using mild detergent and warm water. Blot the probe gently with
soft tissue paper to dry. Avoid excessive rubbing of the glass membrane
and avoid touching it with the fi ngers. Recalibrate after cleaning.

6. TROUBLESHOOTING

Problem

Cause

Solution

No display

Batteries not in place.

a) Insert batteries.
b) Re-insert batteries in correct

polarity.

“LO” displays

in the LCD

Low battery

Replace batteries.

Unstable

reading

a) Electrode not deep

enough in sample
b) Dirty electrode.
c) Broken electrode

a) Place electrode deeper in

sample.
b) Clean electrode and recalibrate.
c) Replace electrode.

Not able to

calibrate

a) Display freezes
b) Faulty electrode
c) Inaccurate buffer

a) Release reading by pressing

HOLD.
b) Replace electrode.
c) Replace expired buffer solutions.

Er1

Buffer value out of tolerance

Use new calibration solution and

recalibrate. Ensure correct pH

buffer group was selected.

Er3

ISE slope not within the

specifi ed tolerance

Recalibrate

Er5

Upon exit of calibration mode,

a 1-point calibration was

attempted with a pH buffer

other than 7.00 or 6.86

Repeat pH calibration using one

or more points which include

either 7.00 (USA) or 6.86 (NIST)

standards.

Oe

Over range: reading exceeds

maximum value

Ensure that the value being

measured is within the range of

the selected mode. Confi rm that

electrode(s) are connected and

working properly.

uE

Under range: reading

exceeds minimum value