LaMotte SALT Testr User Manual

Page 4

background image

3 Bars indicates Battery is full (100%)

2 Bars indicates 50% of the battery life is left
1 Bar indicates 25% of the battery life is left
Blinking battery casing indicates the need to replace
batteries with fresh ones as specified by manufacturer
The sensor electrodes short circuited
Replacement sensor is not connected properly to the
tester during sensor replacement
Measured value or temperature value exceeds the specified
maximum or minimum value
Blinking ‘ATC’, ‘Or’ or ‘Ur’ indicates that there is a shor t
or open circuit at the built in temperature sensor
Calibration error due to temperature value not within the
specified range
Calibration error due to Conductivity/TDS/Salt value not
within the specified calibration standard range

Self-Diagnostic Messages

Er

Er

Er

Er

Er.1

.1

.1

.1

.1

Low battery

indicator

Or / Ur

Or / Ur

Or / Ur

Or / Ur

Or / Ur

(Still)

Over range /
Under range

signal

AAAAATTTTTC / Or /

C / Or /

C / Or /

C / Or /

C / Or /

Ur

Ur

Ur

Ur

Ur (Blinking)

Er

Er

Er

Er

Er.0

.0

.0

.0

.0

Error

Message

Electrode Replacement

When the tester fails to calibrate or gives fluctuating readings in calibration

standards, you need to change the electrode module. You can replace the

electrode module at a fraction of the cost of a new tester.

1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist the

collar counter clockwise (see Figure13-A). Save the ribbed tester collar and

O-ring for later use.

2. Pull the old electrode module away from the tester.

3. Align the four tabs of the new electrode module so that they match the four

slots on the tester (see Figure 13-B).

4. Gently push the module into the slots to sit it in position. Push the smaller

O-ring fully onto the new electrode module. Push the collar over the module

and thread it into place by firmly twisting clockwise.

Note:

Note:

Note:

Note:

Note: It is necessary that you recalibrate the tester prior to measurement after

an electrode replacement.

Accessories

Item

Order Code

ECTESTR, TDSTESTR & SALTESTR

TDSSENSOR

replacement sensor

ECTESTR+ & TDSTESTR+

TDSSENSORPLUS

replacement sensor

Warranty

The waterproof testers are warranted to be free from manufacturing defects for 1 year

and electrode module for 6 months, unless otherwise specified. If repair, adjustment or

replacement is necessary and has not been the result of abuse or misuse within the

time period specified, please return the tester - freight prepaid - and correction will be

made without charge. Out of warranty products will be repaired on a charge basis.

Return of Items

Authorization must be obtained from your distributor before returning items for any

reason.

When applying for authorization, please include information regarding the reason the

item(s) are to be returned.

Note:

Note:

Note:

Note:

Note: Eutech Instruments/Oakton Instruments reserve the right to make improvements

in design, construction and appearance of products without notice. Prices are subject

to change without notice.

Model

Model

Model

Model

Model

ECT

ECT

ECT

ECT

ECTestr11

estr11

estr11

estr11

estr11

ECT

ECT

ECT

ECT

ECTestr11+

estr11+

estr11+

estr11+

estr11+

TDST

TDST

TDST

TDST

TDSTestr11

estr11

estr11

estr11

estr11

TDST

TDST

TDST

TDST

TDSTestr11+

estr11+

estr11+

estr11+

estr11+

SAL

SAL

SAL

SAL

SALTT

TT

TT

TT

TTestr11

estr11

estr11

estr11

estr11

Range: PU

-

0 to 200.0 uS/cm

-

0 to 200.0 ppm

-

LO

0 to 2000 uS/cm

0 to 2000 uS/cm

0 to 2000 ppm

0 to 2000 ppm

-

HI

0 to 20.00 mS/cm

0 to 20.00 mS/cm

0 to 10.00 ppt

0 to 10.00 ppt

0 to 10.00 ppt

Resolution: PU

-

0.1 uS/cm

-

0.1 ppm

-

LO

10 uS/cm

1 uS/cm

10 ppm

1 ppm

-

HI

0.10 mS/cm

0.01 mS/cm

0.10 ppt

0.01 ppm

0.10 ppt

Accuracy

± 1% of Full Scale

Calibration Type

Auto or Manual

Auto or Manual

Manual

Manual

Manual

Calibration Points

1 or 2 points

1, 2 or 3 points

1 or 2 points

1, 2 or 3 points

1 point

Calibration Window

± 50% from each point

Calibration PU

-

2.0 - 200.0 uS/cm

-

2.0 - 200.0 ppm

-

Standard Range:LO 200 - 2000 uS/cm 200 - 2000 uS/cm

200 - 2000 ppm

200 - 2000 ppm

-

(Manual) HI 2.00 - 20.00 mS/cm 2.00 - 20.00 mS/cm

1.00 - 10.00 ppt

1.00 - 10.00 ppt

1.00 - 10.00 ppt

Sensor Type

Two-pin

Cup

Two-pin

Cup

Two-pin

TDS Factor

-

- 0.4 to 1.0 (Default 0.71)-

Temperature

Range in °C

0.0 to 50.0°C

Range in °F

32.0 to 122°F

Resolution

0.1°C (0.1°F)

Accuracy

±0.5°C (±0.9°F)

Calibration point

1 point

Calibration Window

± 5°C (± 9°F)from factory default

ATC

0 to 50°C

Temp Coefficient

2% per °C

Normalization Temp

25.0°C

Auto Off

8.5 minutes after last key press

Operating Temp

0 to 50°C

Power Battery

4 X 1.5V”A76" micro alkaline battery

Battery Life

>150 hrs

LCD Display

Custom Dual Display27mm(H)X21mm(W)

Dimensions

Tester: 16.5 cm X 3.8 cm; 90g

Weight

Boxed: 22cm X 6cmX 5cm; 170 g

Specifications

Changing Batteries

Replace the batteries when the low battery indicator

starts blinking.

1. Open the battery compartment lid (with attached

lanyard loop).

2. Remove old batteries by pulling plastic ribbon.

Replace with fresh ones. Note polarity as shown in

figure 11.

Electrode Maintenance

1. Always keep the sensor electrodes clean. Rinse the electrodes with de-ionized

water and wipe them dry with clean cloth before storing with its protective cap.

For cup type electrodes, remove the white plastic cup and insert to thoroughly

clean viscous solutions. Never scratch electrodes with a hard substance.

2. For better performance, soak the electrode in alcohol for 10 to 15 minutes

and rinse with de-ionized water before starting any measurement process.

This is to remove dirt and oil stains on the electrode which may affect the

accuracy of the measurements.

Figure 13: Removing collar & inserting electrode

68X068056 Rev 2 May 06

+

-

Plastic

Ribbon

B

Rotate ribbed collar

away from you

Ribbed collar

Ribbed collar

Small

O-Ring

Small Tab

Large Tab

Large

O-Ring

Insert Electrode Module

A

Parameter

Factory Default

User calibration (conductivity/TDS/salt)

(Reset)

Temperature unit of measurement

Celsius

(°C)

Temperature offset

0

Auto calibration (for ECTestr11 & ECTester11+)

Enable

1-point calibration

Enable

Conductivity calibration factor (for ECTestr11 & ECTester11+)

1.0

TDS factor (for TDSTestr11 & TDSTester11+)

0.71

Figure 11: Resetting sequence

Reset

Reset option allows you to restore the calibration and other parameters back to

factory default settings.

1. Switch off the tester. Press and hold the HOLD

HOLD

HOLD

HOLD

HOLD key and then switch on the tester

using ON/OFF

ON/OFF

ON/OFF

ON/OFF

ON/OFF key. Release HOLD

HOLD

HOLD

HOLD

HOLD key.

2. The lower display shows ‘rSt’ (reset) and the upper display blinks ‘No’. Use

INC

INC

INC

INC

INC

or DEC

DEC

DEC

DEC

DEC key to

select ‘Yes’ (to proceed with resting) or ‘No’ (to quit without

resetting).

Note:

Note:

Note:

Note:

Note: Press °C/°F

°C/°F

°C/°F

°C/°F

°C/°F key if you wish to skip to measurement mode without making any

selection.

3. Press HOLD

HOLD

HOLD

HOLD

HOLD key to confirm your selection. LCD shows ‘CO’. If ‘Yes’ is selected, the

tester resets to its factory default values as listed below. LCD shows power-up

sequence and tester goes to measurement mode.

Figure 11: Battery compartment

-10-

-11-

ON/OFF

Press & hold HOLD

key and switch on

the tester

YES

Power-up

Sequence

CO

Tester

Off

1

2

2

3

HOLD

HOLD

HOLD

This manual is related to the following products: