Troubleshooting

Problem

Cause

Action

Power on but no
display

Batteries

Insert batteries

Batteries

Verify correct polarity

Batteries

Replace batteries

Unstable readings

Electrode

Immerse electrode more deeply in
sample

Electrode

Condition electrode before first use.
Soak in tap water for 10 minutes.

Electrode

Remove air bubbles caught under
electrode

Electrode

Clean electrode

Electrode

Replace electrode

Static charge

Wiping electrode Rinse electrode and blot. Do not wipe

electrode

Same readings in
different samples

Cracked or broken
electrode

Replace electrode. Avoid plunging
electrode into bottom of container and
stir bars.

Display frozen

HOLD function

Press MODE or turn meter off

Button response

Remove batteries (stored data will be
lost)

Steady “

-1 ” display Wait

Reading not stable yet

Replacing the Electrode
The TRACER is shipped with an electrode attached. If the electrode
needs to be replaced, follow these steps for removing and connecting
electrodes.
1. Turn meter off.
2. To remove the electrode, turn the collar counter-clockwise and

remove it.

3. Gently rock the electrode from side to side, pulling it downwards,

until it disconnects from the meter.

4. To attach an electrode, carefully plug the electrode into the meter

socket. Note that the electrode connector is keyed to ensure a
proper connection.

5. Tighten the electrode collar firmly enough to make a good seal. A

rubber gasket will seal the electrode with the meter.

9