LaMotte Tracer pH Meter User Manual
Page 14
14
Problem
Cause
Action
Erratic LCD
display
Samples have low ionic
strength (lack salt);
e.g. distilled, deionized,
boiled, lake water (high
pressure)
For each 50 mL of sample add
1 drop (50 mL) of saturated
Potassium Chloride (KCl). No
alteration in pH will occur by
inert KCl.
Unexpected
readings
Buffers
Calibrate with fresh buffers
Buffers
Calibrate with buffers that
bracket sample pH
Display frozen
HOLD function
Press MODE/HOLD or turn
meter off
Button response
Remove batteries press ON/OFF
button for 3 seconds. Reinsert
batteries (stored data will be
lost)
Steady “-1 ”
display
Wait
Reading not stable yet
EXPAND YOUR TRACER
Interchangeable electrodes are available to convert the pH TRACER to a Total
Chlorine TRACER or an ORP TRACER.
Remember to ask for instructions and appropriate reagents or buffer tablets
when ordering the Total Chlorine or ORP electrodes.