beautypg.com

LaMotte Tracer pH Meter User Manual

Page 14

background image

14

Problem

Cause

Action

Erratic LCD
display

Samples have low ionic
strength (lack salt);
e.g. distilled, deionized,
boiled, lake water (high
pressure)

For each 50 mL of sample add
1 drop (50 mL) of saturated
Potassium Chloride (KCl). No
alteration in pH will occur by
inert KCl.

Unexpected
readings

Buffers

Calibrate with fresh buffers

Buffers

Calibrate with buffers that
bracket sample pH

Display frozen

HOLD function

Press MODE/HOLD or turn
meter off

Button response

Remove batteries press ON/OFF
button for 3 seconds. Reinsert
batteries (stored data will be
lost)

Steady “-1 ”
display

Wait

Reading not stable yet

EXPAND YOUR TRACER

Interchangeable electrodes are available to convert the pH TRACER to a Total
Chlorine TRACER or an ORP TRACER.

Remember to ask for instructions and appropriate reagents or buffer tablets
when ordering the Total Chlorine or ORP electrodes.

Total Chlorine TRACER Electrode,
0.0-10.0 ppm

Code 1732

The Total Chlorine TRACER
Electrode (Code 1732) requires the
use of TRACER TCL Tablets. Order
Code 7044-J (100 pack).

ORP TRACER Electrode, ±999 mV

Code 1734

The ORP TRACER Electrode (Code
1734) requires an initial soaking in a
pH 4.0 buffer solution. Order pH 4.0
Mini Buffer Tablets/100 pack (Code
3893-J).