MANGANESE TEST PROCEDURE - PAN METHOD

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

MANGANESE

4.

Push the

READ button

to turn the meter

on. Press the

ZERO

button

and hold it for 2

seconds until

bla

is displayed.

Release the button

to take a blank

reading (0.0 ppm).

I200

COLORIMETER

• • • • • • • • • • • • • • • • • •

READ

ZERO

bla

Insert the tube into

the chamber, being

3.

Rinse and fill a

colorimeter tube

(0290) to the
10 mL line with

sample water. Cap

and wipe dry.

2.

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

1.

Use the 1.0 mL

pipet (0354) to

add 2.0 mL of

*Hardness Buffer

Reagent (4255).

5.

6.

Swirl the tube to mix.

8.

7.

Add two drops of

*Sodium Cyanide,

10% (6565).

Swirl the tube to mix.

sure to align

the index line

with the arrow

on the meter.

Close the lid.

This tube is

the blank or

zero.