LaMotte 1200 COLORIMETER IRON 1200-FE User Manual

Page 2

background image

IRON TEST PROCEDURE - 1,10 PHENANTHROLINE METHOD

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

IRON

2

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

Rinse and fill a

colorimeter tube

(0290) to the 10

mL line with

sample water.

Cap and wipe

dry.

5.

3.

1.

2.

4.

6.

I200

COLORIMETER

• • • • • • • • • • • • • • • • • •

READ

bla

ZERO

Cap and invert the

tube 15 times to

mix.

Push the READ

button to turn the

meter on. Press the

ZERO

button

and hold it for 2

seconds until

bla

is

displayed. Release

the button to take a

zero reading
(0.00 ppm).

Remove tube from

colorimeter. Use the

0.1 g spoon (0699) to

add one measure of

*Iron Reducing

Reagent (2777).

8.

7.

Remove the cap and

add 6 drops of *Acid

Phenanthroline

Indicator (2776).

Cap and invert the

tube 3 times to mix

reagents.

L Mott