HF scientific DRT-15CE Portable Turbidimeter User Manual
Page 14

DRT-15CE (5/10)
Page 6
Each kit contains:
--
Instruction manual
--
1 liter of 4000 NTU Stock Suspension
--
1 Gallon (3.79 liters) turbidity-free water
--
4 Sample cuvettes (28 mm)
--
4 Light Shield Caps
--
Graduated Pipettes 1 ea. in 1 ml, 10 ml, & 25 ml
--
1 Calibration Standard
The following table gives the recommended dilutions of the stock suspension.
Be sure to adequately mix the stock suspension prior to removing a portion
for dilution.
PROCEDURE
NTU Value
Pipette
Amount
Base
Formazin
Volumetric
Flask
10.0 NTU
10.0ml
2.5ml
of 4000 NTU
1000ml
100.0 NTU
10.0ml
5.0ml
of 4000 NTU
200ml
1000.0 NTU
25.0ml
25.0ml
of 4000 NTU
100ml
NOTE:
When the prepared samples start to flocculate, they are
unreliable and fresh ones must be made. This will occur
more rapidly for the lower value diluted suspen sions.
Prepare 40 NTU and lower standards daily.
7.2
ELECTRONIC CALIBRATION USING FACTORY
CERTIFIED PROCAL STANDARDS (CAT. NO. 39071)
The DRT Turbidimeters have been carefully calibrated at the factory.
However, should the Electronic P.C. Board, the Photo Detec tors, or the Light
Source be replaced recalibration may be easily accomplished at your facility.
To carry out a complete calibration the following ProCal Standard values are
required: