7.0

CALIBRATION PROCEDURES

7.1

Calibration Standards

7.1.1 ProCal Standard Set (optional) Catalog No. 39071

HF ProCal Standards are recommended and certified by HF scientific.

They are traceable to freshly prepared formazin primary standards. These

standards are very easy to use off the shelf anytime without preparation

making them an ideal turbidity standard. A Certificate of Traceability is

available on request to the HF scientific Customer Service Department. HF

ProCal Standards may be used for calibration of HF turbidimeters. Order

from HF scientific, inc.

NOTE: Do not freeze standards.

Do not leave standards in the measuring well for extended

periods.

Do not shake standards.

Each ProCal Standard Kit contains:

--

0.02 Calibration Standard

--

Certified ProCal Standards 10.00, 100.0, 1000 NTU

Standards are contained in preselected cuvettes with light

shield caps.

--

A sturdy storage case

7.1.2 Standard Formazin Solutions

Calibration of this instrument is based on Formazin, a material which is

made by polymerization.

Calibration samples may be obtained by diluting Formazin stock suspension

using “Turbidity-Free” water. Formazin stock suspension can be prepared

by the user (Reference Standard Methods For Examination of Water and

Wastewater) or a kit can be purchased from, HF scientific, inc.,

Catalog No.

50040.