beautypg.com

Gossen MAVOSPEC User Manual

Page 18

background image

GOSSEN Foto- und Lichtmesstechnik GmbH

18

3

Trigger Input / Output

MAVOSPEC is equipped with input and output signal lines dedicated for starting
measurement from external device and triggering external devices when measurement
is started.

Trigger In line - high state on this line causes the same effect as pressing button on the
device.

Trigger Out line - at the measurement start MAVOSPEC sets high state on this line.
The high state is maintained throughout the period of measurement.

Trigger lines state:

Trigger Technical Data

Connector

Minijack 3.5mm, stereo, 3-pin

Trigger in line

Input line, type open collector with galvanic
isolation. Voltage 4.5 ... 9 V. Min current 1 mA

Trigger in signal

Recommended high state >= 50 µs,

Trigger out line

Output line, type open collector with galvanic
isolation. Voltage 3 ... 9 V.
Recommended current <= 1 mA