8 Technical specifications

890 Titrando

■■■■■■■■

35

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)

Measuring mode
Upol

Determination with adjustable polarization voltage

Polarization
voltage

–1,225 - +1,225 mV (increment: 25 mV)
–1,250 - +1,250 mV: non-guaranteed values, dependent on reference
voltage +2.5 V

Measuring
range

–120 - +120 µA

Resolution

0.1 µA

8.2

Internal dosing device

Cylinder volume of
the exchange unit

1 mL, 5 mL, 10 mL, 20 mL or 50 mL

Resolution

20,000 steps per cylinder volume

8.3

Power connection

Supply voltage

100 - 240 V

Frequency

50 - 60 Hz

Power consump-
tion

Maximum 45 W

Fuse

Electronic overload protection

8.4

Safety specifications

Design and testing

EN/IEC 61010-1

UL 61010-1

CSA-C22.2 No. 61010-1

Protection class I

Safety instructions

The documentation contains safety instructions which have to be fol-
lowed by the user in order to ensure safe operation of the instrument.