Application Bulletin 320 e
Installation Instructions of TitrIC 6
Page 5/16

2 Installation

The following is a detailed description of the TitrIC 6 System's installation.

We strongly recommend that the individual steps are carried out in the order given below!

2.1 815 Robotic USB Sample Processor XL

The sample processor is placed at the side at which it is most easily accessible. Please note that the
following procedure refers exclusively to the version in which the sample processor is located at the left-
hand side of the system.

ƒ On the right-hand side of the 815 Robotic USB Sample Processor XL, attach the stand

support (6.2001.070) without the horizontal brace to the tower of the 815 (hexagon screw
beneath membrane pump 1). The brace will be installed later.

ƒ Mount both Swing Heads (2.786.0010) onto the 815 (see installation instructions 2.2.7, p 19).

ƒ Do not screw the swing arm (6.1462.050) nor the titration arm onto the Swing Head!

ƒ The

Sample Rack supplied is screwed onto the 815.

ƒ In the vicinity of the 815 the two canisters (standard equipment of 815) are placed on the floor:

One 10 L canister is used as a waste container, the second one as a reservoir for deionized
(DI) water used for rinsing the external cell.

ƒ From the smaller opening of the canisters containing DI water, an M8 PTFE tubing connection

of suitable length is led to the connection box made of white plastic of membrane pump 1 /
tower 1 and connected.

ƒ From the front opening of the waste canister an M8 PTFE tubing connection of suitable length

is led directly to membrane pump 1 / tower 2 (black plastic) and connected.

A more detailed description is given in the instructions for use of the 815 Robotic USB Sample
Processor XL.

2.2 856 Conductometer (2.856.0010)

The 856 Conductometer is placed directly at the right of the sample processor. For detailed information
about the usage check the Instructions of Use of the 856.
Before measurements it is recommended to carry out the calibration of the cell constant in tiamo

TM

. For

further information consult the Installation Instructions of the 856 Conductometer.

2.3 Titrando

(2.905.0010)

Directly to the right of the 856 Conductometer, place the Titrando with the two bottles of the titration
solution (0.1 mol/L HCl, at the rear, with 20 mL Dosing Unit and 800 Dosino) and the acidification
solution (0.1 mol/L HNO

3

, at the front, with eluent bottle attachment).

2.4 Dosino holder (6.2055.100)

The Dosino holder (6.2055.100) is placed behind the Titrando and the 856 Conductometer; the 50 mL
(6.3032.250) and 10 mL (6.3032.210) Dosing Units are screwed onto this holder at a later stage.

2.5 850 Professional IC AnCat – MCS (2.850.3030)

Place the 850 Professional IC AnCat - MCS (2.850.3030) to the right of the Titrando. The IC
instruments can already be installed. Please consult the Instructions for Use for the Professional IC
instruments.