Application Bulletin 320 e
Installation Instructions of TitrIC 6
Page 11/16
6.2 Configuration
of
tiamo
TM
Æ The Controller cable (6.2151.000), which is already connected to the 815, is now connected to an
USB interface on the PC. Wait a few seconds. The driver installation for 905 and 815 follows. It’s not
necessary to start windows update, click on
automatic installation of the software
for both.
Æ The controller cable (6.2151.000), which is already connected to the 856 conductivity module, is now
connected to an USB interface on the PC
Æ tiamo
TM
2.1 is started
6.2.1 Devices
Connected USB devices are automatically recognized when tiamo
TM
is started. After confirmation of the
automatically generated request the devices are stored in the configuration. Keep the proposed device
names.
6.2.2 Swing Head configuration
Under
Configuration
at the top left, double-click on
815_1
under
Devices
, click on the entry
Tower 1
and
afterwards on
Tower 2
, set the
Axial distance
to 196.0 mm and in the middle of the window under
Swing
Head
click on
Configuration
and
Yes
and then enter the following values:
Configuration of the Swing Head at tower 1 Configuration of the Swing Head at tower 2
Confirm the entries twice with <
OK>
and then mark
815_1
under
Devices
and click on
Edit /
Initialize
. Wait a
few seconds. The titration arm (6.1462.050) is now mounted onto the Swing Head at tower 1, the DIS-
COVER arm (6.1462.080) onto the Swing Head at tower 2 and is connected to
Ext. Pump 2
of tower 2.
6.2.3 Special beaker definition
Under
Configuration
mark
815_1
, select
Edit / Properties / Rack / Rack data / Special beaker
,
Special beaker 1
, click
on
Edit
and define the position to be used for rinsing as
Rack position
. The last position is normally used
as the rinsing position. A sample vessel containing DI water should always stand in this position.
Confirm by clicking on
twice. No cap should be put on the DI water in the provided methods.
6.2.4 Lift
positions
Under
Configuration / Tools / Manual operation
click on
815 Sample Changer / Tower 1
in the left-hand column
and use the entries
Move
and
Assign position
(See 815 Instructions for Use, chapter 2.4.6, page 47) to
define suitable values (click on
each time) for:
Shift position
for changing the rack position (approx. 53 mm)
Work position
for aspirating of the sample (approx. 173 mm)
Special position
for dipping for the conductivity measurement (approx. 130 mm)
For the special position, you must define the work position again, then at the right beside
Work position
click on
Tower
and select
Special position 1
.
Please repeat these steps to define the settings for tower 2.