Metrohm 803 Ti Stand User Manual
Page 22
![background image](/manuals/587038/22/background.png)
3.4 Mounting the accessories
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
803 Ti Stand
5 Fasten the one end of the aspiration tubing to the tubing adapter,
the other to the connection nipple for the waste on the rear of the
titration stand.
6 Insert the 6.1819.050 cannula through the 6.1446.090 stopper into
the waste bottle.
7 Screw the 6.1805.200 PTFE tubing into the 6.1446.090 stopper.
8 Screw the 6.1543.120 aspiration tip to the other end of the PTFE
tubing.
9 Insert the aspiration tip down to the bottom of the titration vessel to
be emptied and fasten it to the titration vessel lid.
Note
The waste bottle should periodically be emptied.
Mounting the equipment for aspirating solvent
Mount the solvent bottle as follows:
1 Fill the 6.1609.010 adsorber tube and the 6.2811.000 molecular
sieve with cotton.
2 Use the other half of the 6.1801.120 PVC tubing previously cut in
two pieces for aspirating the solvent. Fasten the one end of the tub-
ing to the lower end of the adsorber tube, the other end to the con-
nection nipple for aspirating the solvent.
3 Equip the 6.1602.105 bottle attachment with the 6.1446.040 threa-
ded stopper and the 6.1609.010 adsorber tube.
4 Secure the adsorber tube with the 6.2023.020 SGJ clip.
5 Insert the 6.1819.030 cannula into the last open hole (with medium-
sized diameter) of the 6.1602.105 bottle attachment.
6 Screw the 6.1602.105 bottle attachment onto the solvent bottle.