7 Appendix
806 Exchange Unit
■■■■■■■■
31
Table 1
Typical measurement deviation of Metrohm exchange units
Cylinder volume
max. systematic deviation
1 mL
± 3 µL
5 mL
± 15 µL
10 mL
± 20 µL
20 mL
± 30 µL
50 mL
± 50 µL
7.2.2
The ISO/EN/DIN standard 8655-3
The Metrohm exchange units fulfil the requirements of the ISO/EN/DIN
standard 8655-3 Volume measurement instruments with pistons –
Part 3: Piston burets. Metrohm guarantees that its exchange units are
in compliance with the following limit values at the time of shipment:
Table 2
Permissible limit values as per ISO/EN/DIN 8655-3
Cylinder
volume
max. systematic
measurement deviation
max. permissible
measurement deviation
1 mL
± 0.6 %
± 6 µL
± 0.1 %
± 1 µL
5 mL
± 0.3 %
± 15 µL
± 0.1 %
± 5 µL
10 mL
± 0.2 %
± 20 µL
± 0.07 %
± 7 µL
20 mL
± 0.2 %
± 40 µL
± 0.07 %
± 14 µL
50 mL
± 0.2 %
± 100 µL
± 0.05 %
± 25 µL
The Metrohm agents worldwide offer the possibility of on-site exchange
unit inspections and certifications with respect to accuracy. We recom-
mend that an accuracy inspection be performed when the dosing cylin-
ders and dosing pistons of an exchange unit are replaced.