4 Installation

940 Professional IC Vario ONE/SeS/Prep 3 (2.940.1430)

■■■■■■■■

37

Start manual control in MagIC Net and select the high-pressure
pump: Manual

Manual control Pump

Flow: in accordance with column leaflet
On

Rinse the guard column with eluent for approx. 5 minutes.

Use manual control in MagIC Net to stop the high-pressure pump
again: Off.

4.22

Connecting the separation column

CAUTION

New separation columns are filled with solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow information from
the manufacturer).

NOTE

The separation column may not be connected until after the instrument
has already been put into operation once (see Chapter 4.20, page 33).
The guard column and separation column have to be replaced by a
coupling (6.2744.040) until then.