4 Installation

940 Professional IC Vario ONE/SeS/Prep 3 (2.940.1430)

■■■■■■■■

31

Connecting the capillaries of the SMP

NOTE

Pump tubing connections with a filter (6.2744.180) have to be installed
on the outlet of the peristaltic pump's pump tubing in order to protect
the SPM from foreign particles and bacterial growth.

CAUTION

As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.

Squeezed capillary ends can be shortened with the capillary cutter
(6.2621.080).

Connect the SPM capil-
lary …

… to …

… using this accessory

in

Sample Processor

or

(see Sample Processor
manual)

or

a 6-port valve used for
measuring the sample
quantity.

6.2744.010

out

Sample input of the injec-
tion valve

6.2744.010

rinsing solution

Pump tubing output for
rinsing solution

Pressure screw
(6.2744.070)

regenerant

Pump tubing output for
regeneration solution

Pressure screw
(6.2744.070)

waste rins.

Waste container

Directly or via waste collec-
tor (6.5336.000)

waste reg.

Waste container

Directly or via waste collec-
tor (6.5336.000)