Dosing unit at 800 dosino (2) and 800 dosino (3), Dosing unit at 800 dosino (4), Solutions – Metrohm MVA-20 User Manual
Page 6: Devices – 919 ic autosampler plus for cvs, Tower

Application Bulletin 402/1 e
Installation instruction for MVA-20
Dosing rate port 1
Maximum
mL/L
Dosing rate port 2
Maximum
mL/L
Dosing rate Fill port
Maximum
mL/L
Dosing rate special port
Maximum
mL/L
Tubing parameters
Port
Length
Diameter
Dosing port 1
Port 1
138 cm
2 mm
Dosing port 2
Port 3
0 cm
2 mm
Fill port
Port 2
25 cm
2 mm
Special port
Port 4
0 cm
2 mm
Valve disk
Rotating direction
automatic
3.3.2. Dosing unit at 800 Dosino (2) and 800 Dosino (3)
Dosing Unit name as used in the viva method templates.
800 Dosino (MSB 2)
894/D2 2mL Brightener
800 Dosino (MSB 3)
894/D3 2mL Suppressor
Parameters for preparation
Dosing port Prep/empty
Dosing port 2
Dosing rate port 1
2.0
mL/L
Dosing rate port 2
Maximum
mL/L
Dosing rate Fill port
Maximum
mL/L
Dosing rate special port
Maximum
mL/L
Tubing parameters
Port
Length
Diameter
Dosing port 1
Port 1
80 cm
0.3 mm
Dosing port 2
Port 3
0 cm
2 mm
Fill port
Port 2
55 cm
2 mm
Special port
Port 4
0 cm
2 mm
Valve disk
Rotating direction
automatic
3.3.3. Dosing unit at 800 Dosino (4)
Dosing Unit name as used in the viva method templates.
800 Dosino (MSB 4)
894/D4 2mL Std or sample
Parameters for preparation
Dosing port Prep/empty
Dosing port 2
Dosing rate port 1
2.0
mL/L
Dosing rate port 2
maximum
mL/L
Dosing rate Fill port
4.0
mL/L
Dosing rate special port
maximum
mL/L
Tubing parameters
Port
Length
Diameter
Dosing port 1
Port 1
80 cm
0.3 mm
Dosing port 2
Port 3
0 cm
2 mm
Fill port
Port 2
201 cm
1 mm
Special port
Port 4
0 cm
2 mm
Valve disk
Rotating direction
automatic
3.4.
Solutions
Solutions that should automatically be dosed have to be
defined in the viva «Configuration» and need to be
assigned to the Dosing Unit which is used for the dosing.
The following table shows the solution names and assigned
Dosing Units as used in the viva method templates.
Solution name
Dosing Unit
VMS
894/D1 50mL VMS
Brightener concentrate
894/D2 2mL Brightener
Suppressor concentrate
894/D3 2mL Suppressor
Std or sample
894/D4 2mL Std or sample
S
OLUTION TYPE
for «VMS» and «Suppressor concentrate»
has to be A
UXILIARY SOLUTION
, for «Brightener concentrate»
and «Std or sample» it is S
TANDARD SOLUTION
.
For a step by step description please see the Tutorial
chapter «5.1.4 Define solutions».
3.5.
Devices – 919 IC Autosampler plus for CVS
The «919 IC Autosampler plus for CVS» is automatically
recognized by the viva software. When an instrument is
connected for the first time it needs to be entered in the viva
«Configuration». The corresponding dialogue will pop up
automatically.
Device name*
919 IC Autosampler plus 1
* as used in the example method
For a step by step description, using the example of an 858
Professional Sample Processor, please see the Tutorial
chapter «6.1.1 Configuring the instrument».
3.5.1. Tower
Tower parameters
Max. stroke path
133
mm
Min. beaker radius
Off
Lift rate
25 mm/s
mm/s
Axial distance
189.0
mm
Page 6 of 12