Substances, Standards, Calibration – Metrohm MVA-20 User Manual
Page 10: Results, Operation, Suppressor determination series, Manual operation, Workplace

Application Bulletin 402/1 e
Installation instruction for MVA-20
4.2.1. Substances
In the «Substances» part settings for peak recognition and
baseline parameters are defined.
4.2.2. Standards
In the «Standards» part the concentration of the used
standard solution is defined.
4.2.3. Calibration
In the «Calibration» part the calibration method, such as DT
or MLAT, is defined as well as the regression type.
4.2.4. Results
In the «Results» part on the tab «Results» the substance is
selected for which the concentration should be calculated
and displayed.
5. Operation
5.1. Suppressor determination series
5.1.1. Manual operation
•
Dosing Unit «894/D1 50mL VMS» connected to Dosino
1 has to be prepared with VMS.
•
Suppressor standard solutions and plating bath
samples are placed on position 1 to 27 of the sample
rack using the sample vials for 50 mL (6.2747.010).
•
On the position after the last sample a vial with
deionized water is placed to rinse Dosino 4 after the
end of the suppressor sample series.
Fig. 14: Example positions of solutions on the sample rack
6.2041.490 for suppressor determination
5.1.2. Workplace
•
In the «Run» window on the tab «Determination series»
create a sample table
Fig. 15: Example sample table for suppressor determination
•
To run a calibration with a standard solution select
sample type S
TANDARD
. For a determination the sample
type S
AMPLE
has to be selected.
•
No «Sample amount» needs to be defined in the
sample table, since addition volumes for the standard
and the sample are defined in the method commands
ADD
STD and ADD
SAMPLE
DT.
Page 10 of 12