1 Introduction
906 Titrando
■■■■■■■■
3
1.3
Titration modes – Measuring modes – Dosing com-
mands
The 906 Titrando supports the following titration modes, measuring
modes and dosing commands:
■
DET
Dynamic equivalence point titration. The reagent addition is carried out
in variable volume steps.
Measuring modes:
– pH (pH measurement)
– U (potentiometric voltage measurement)
– Ipol (voltametric measurement with selectable polarization cur-
rent)
– Upol (amperometric measurement with selectable polarization
voltage)
■
MET
Monotonic equivalence point titration. The reagent addition is carried
out in constant volume steps.
Measuring modes:
– pH (pH measurement)
– U (potentiometric voltage measurement)
– Ipol (voltametric measurement with selectable polarization cur-
rent)
– Upol (amperometric measurement with selectable polarization
voltage)
■
SET
Endpoint titration at one or two specified endpoints.
Measuring modes:
– pH (pH measurement)
– U (potentiometric voltage measurement)
– Ipol (voltametric measurement with selectable polarization cur-
rent)
– Upol (amperometric measurement with selectable polarization
voltage)
■
STAT
Endpoint titration during which the measured value is kept constant.
Measuring modes:
– pH (pH measurement)
– U (potentiometric voltage measurement)