Metrohm 930 Compact IC Flex Oven/ChS/PP User Manual

Page 36

background image

4.15 Initial start-up

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

930 Compact IC Flex Oven/ChS/PP (2.930.2300)

4 Deaerating the high-pressure pump

Use the syringe to aspirate eluent until there are no more air

bubbles in the eluent.

5 Setting the contact pressure of the peristaltic pump

Set the contact pressure of the peristaltic pump (see "Setting the
contact pressure correctly", page 23)
.

6 Rinsing the instrument without columns

Rinse the instrument (without columns) with eluent for 5 minutes.