7.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
58
■■■■■■■■
872 Extension Module IC Module
7 Troubleshooting
7.1
Problems and their solutions
Problem
Cause
Remedy
Marked rise in pres-
sure
6.2821.120 inline filter
blocked.
Replace 6.2821.130 filter (see Chapter 6.5,
page 53).
Injection valve – valve
blocked.
Have the valve cleaned (by Metrohm service
technicians).
Drift of the baseline
Eluent – Evaporation of the
organic solvent in the elu-
ent.
Check the eluent bottle attachment (see Figure
14, page 25).
Peak areas lower
than expected
Sample – leak in the sam-
ple path.
Check the sample path.
Sample – blockage in the
sample path.
Check the sample path.
Sample – sample loop not
(completely) filled.
Prolong the sample transfer time.
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves (see Chapter 6.4.2, page
43).
Eluent – Leak in the eluent
path.
Check the eluent path
Eluent – Blockage in the
eluent path.
Check the eluent path
High pressure pump –
defective piston seals.
Replace piston seals (see Chapter 6.4.2, page
43).
Pulsation damper not con-
nected.
Connect pulsation damper (see Chapter 4.8,
page 34).
Background conduc-
tivity too high
Incorrect eluent.
Change eluent (see Chapter 6.3.2.3, page 42).
Individual peaks
greater than expec-
ted
Sample – carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path